Description: | DQR No: Platform: sgpsiros13
Subject: SIROS data corrupted when tower was down
Date Submitted: 12/27/94
Submitted By: Marvin L. Wesely ___ Instrument Mentor
___ EST Member
___ Science Team Member
_x_ Other _Instrment Team leader_
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement:
What level data: (raw,a0,a1,b1,c1 etc):
all data levels for sgpsiros13
What location was the data collected at: SIROS CF
Period of time in question
Begin Date 18/AUG/94 Time 15:30 (GMT)
End Date 18/AUG/94 Time 16:30 (GMT)
Data should be labeled:
_x_ questionable _x_ All data fields affected
___ incorrect ___ Only some data fields affected
___ wrong calibration
___ others
Discussion of Problem:
The 10-m tower at the central facility central cluster was lowered
by site operations personnel and work was carried out on the SIROS
cabling and possibly the MFRSR-supplied data logger. Some data
might be missing, some might be out of range, and some might appear
reasonable but should not be trusted for the time interval and
location specified.
Other observations/measurements impacted by this problem:
SMOS sensors when tower was down between 1420 and 1520 GMT; see
DQR P940822.1
MFR10 sensor when tower was down between 1420 and 1520 GMT, as noted in
a separate DQR
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)
Flat data with this comment.
Data Processing Notes Date |