DQRID : D961008.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/18/1996000010/03/19962359Suspect
more
Subject:
SGP/MFR10M/C1 - Corrupt or truncated data
DataStreams:sgpmfr10mC1.b1, sgpmfr10mC1.a0
Description:
On about 96/08/18, the data logger board for the downlooking 10m MFR was modified to test 
an upgrade.  Unfortunately this upgrade did not work as planned, and some data was either 
corrupted or truncated.  You can't miss the corrupted data (from which you can infer 
albedos much greater than one!).  The truncated data is truncated in time, by which we mean 
that only a half day's worth of data was collected.  The daily time span of these 
truncated data is from midnight to about 1 PM (local time). At about 96/10/03, these problems 
were corrected.

Note:  The "begin date" given for this DQR may be off a day or two, so beware.
Suggestions: 
Measurements:sgpmfr10mC1.b1:
  • up_hemisp_broadband
  • up_hemisp_narrowband_filter6
  • up_hemisp_narrowband_filter4
  • up_hemisp_narrowband_filter5
  • cosine_solar_zenith_angle
  • callang_flags
  • head_temp
  • up_hemisp_narrowband_filter2
  • up_hemisp_narrowband_filter3
  • up_hemisp_narrowband_filter1
more
sgpmfr10mC1.a0:
  • up_hemisp_narrowband_filter1
  • up_hemisp_broadband
  • up_hemisp_narrowband_filter3
  • up_hemisp_narrowband_filter2
  • up_hemisp_narrowband_filter5
  • up_hemisp_narrowband_filter4
  • up_hemisp_narrowband_filter6
  • logger_volt
  • callang_flags
  • head_temp
more