Description: | IRT was out of service for calibration check. MWR was powered
down on 96/12/03 1937-1949 to remove IRT and on 96/12/12 1848-1920
to re-install IRT.
The following is from the NREL Metrology Laboratory test report of 12/10/97:
Temperature Nominal Value Measured Value
----------- ------------- --------------
0 C 273.2 K 274.1 K
10 283.2 283.6
20 293.2 293.6
30 303.2 303.2
40 313.2 313.2
In the range tested, the temperature difference is within the
resolution of the instrument, so the IRT was not adjusted and
the calibration factor was not changed. |