DQRID : D970213.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/21/1995000002/04/19971912Suspect
more
Subject:
Reprocess: Wrong calibration factor applied to NIP
DataStreams:sgpsirosE10.a1
Description:
From about 950721 to about 970204 (19:12 GMT) the wrong calibration
factor was applied to the NIP at site E10.  The calibration factor
that was used was 1.00E-06 Volts/(W m^2).  The correct calibration
factor is 8.338E-06 Volts/(W m^2).  Thus the calibrated NIP values
found in the netCDF files are too large by a factor of 8.338.
Suggestions: 
Divide the calibrated NIP values given in the netCDF files by 8.338.
Measurements:sgpsirosE10.a1:
  • therm_volt
  • lat
  • flag_zero_cosine1-24
  • direct_norm_broadband
  • up_short_hemisp
  • logger_volt
  • alt
  • flag_zero_divisor1-24
  • up_long_dome_temp
  • down_long_dome_temp
  • base_time
  • short_direct_normal
  • mfrsr_temp
  • diffuse_hemisp_narrowband
  • up_long_hemisp
  • flag_ln_error1-24
  • hemisp_broadband
  • lon
  • channel
  • direct_norm_narrowband
  • down_short_hemisp
  • hemisp_narrowband
  • flag_nighttime1-24
  • up_long_case_temp
  • down_long_case_temp
  • down_long_diffuse_hemisp
  • logger_temp
  • diffuse_hemisp_broadband
  • time_offset
  • down_short_diffuse_hemisp
more