Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/05/1997 | 0000 | 05/05/1997 | 2359 | Does not affect quality |
Subject: | SGP/MWR/C1 - IRT Calibration |
DataStreams: | sgpmwrlosC1.b1, sgpmwrlosC1.a1 |
Description: | Comparison of the uplooking IRT with AERI for 5 May 1997 during a time when the 9-11 micrometer sky temperature was in the range 220-228 K, according to AERI, the IRT indicated approximately 3 K higher. This is outside of the specified uncertainty in the IRT and indicates that the calibration needs to be revised. |
Suggestions: | The instrument calibration must be checked. |
Measurements: | sgpmwrlosC1.a1: more
sgpmwrlosC1.b1: more |