Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/27/1997 | 0530 | 12/18/1997 | 1805 | Incorrect |
Subject: | SGP/EBBR/E12 - AEM Failure and Replacement |
DataStreams: | sgp5ebbrE12.b1, sgp30ebbrE12.b1, sgp15ebbrE12.b1 |
Description: | The E12 AEM did not exchange for most of the period above, yielding small negative home signals. The AEM was replaced on 18 December by site operations personnel. All 30 minute e, h, and bowen data are incorrect for the times when the home signal is incorrect, and the temperature, relative humidity, and vapor pressures do not represent properly switched values. |
Suggestions: | |
Measurements: | sgp15ebbrE12.b1: more
sgp5ebbrE12.b1: more sgp30ebbrE12.b1: more |