DQRID : D980126.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/19/1998113301/19/19981133Incorrect
more
Subject:
SGP/SONDE/C1  - Incorrect surface value
DataStreams:sgpsondewrpnC1.a1, sgpsondewrpnC1.a0, sgpsondewrpnC1.00
Description:
Incorrect values of surface temperature and relative humidity were entered by the 
operator.  The surface temperature was entered as 17.5 degC; the first value reported by the 
sonde was -3.6 degC.  The surface humidity was entered as 26; the first value reported by the 
sonde was 81%. It appears that the operator used the inside temperature/humidity rather 
than the sonde-reported outside temperature/humidity.
Suggestions: 
Do not use data
Measurements:sgpsondewrpnC1.a0:
  • tdry
  • dp
  • rh
more
sgpsondewrpnC1.a1:
  • rh
  • dp
  • tdry
more
sgpsondewrpnC1.00:
  • Raw data stream - documentation not supported
more