DQRID : D990915.142
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/30/1998193004/30/19982330Incorrect
more
Subject:
SHB/NIMFR/C1 - Operations problem
DataStreams:shbnimfrC1.00, shbnimfrC1.a1, shbnimfrC1.a0, shbnimfrC1.b1
Description:
This information is automatically generated from a data base. There may be errors in the 
measurement names or measurement names for certain data streams may not as yet have been 
loaded into the data base.
	 Observed a sudden jump	in data; this jump is completely uncorrelated to other data, e.g. 
the ceilometer and the dabul show a constant cloud cover from	15:00-21:00; However, the 
voltage and detector temperature jumps synchronously with the	signal,	so there must have 
been	problems with the instrument at	that time.
Suggestions: 