Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/11/1999 | 0000 | 06/29/1999 | 2145 | Incorrect |
Subject: | SGP/EBBR/E26 - Left T/RH Probe Temperature Data Incorrect |
DataStreams: | sgp5ebbrE26.b1, sgp15ebbrE26.b1, sgp30ebbrE26.b1 |
Description: | Left T/RH probe temperature is frequently offscale during the period. Examine the data (compare with the the T/RH right temperature) to determine which times are incorrect. Replacement of the probe on 29 June 1999 solved the problem. Affected data are: 5 minute: tair_bot, tair_top 15 minute: tair_l 30 minute: tair_bot, tair_top, e, h, bowen |
Suggestions: | |
Measurements: | sgp5ebbrE26.b1: more
sgp15ebbrE26.b1: more sgp30ebbrE26.b1: more |