Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/02/1999 | 0400 | 02/23/1999 | 1500 | Incorrect |
Subject: | SGP/EBBR/E15 - Temperatures Incorrect |
DataStreams: | sgp5ebbrE15.b1, sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | Poor contact of the reference temperature probe leads to the J-panel or multiplexer caused frequent offscale and incorrect values of the reference temperature. All other temperature measurements (air, soil, and humidity probe) depend on the reference temperature for a correction. Therefore all temperature measurements are frequently offscale and incorrect during the specified period. The data clearly shows the offscale times. |
Suggestions: | |
Measurements: | sgp5ebbrE15.b1: more
sgp15ebbrE15.b1: more sgp30ebbrE15.b1: more |