Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/25/1998 | 1700 | 03/10/1999 | 1600 | Incorrect |
Subject: | SGP/EBBR/E20 - Right Relative Humidity Offset Low |
DataStreams: | sgp15ebbrE20.b1 |
Description: | The right relative humidity has been offset low approximately 5% since the installation of the recalibrated EBBR system on 25 Nov. 1998. This only affects the absolute value of the relative humidities used in the flux calculations; the fluxes themselves are not affected because of the removal of offsets by the AEM switching. The right temperature/humidity probe was replaced on 10 March 1999. |
Suggestions: | |
Measurements: | sgp15ebbrE20.b1: more
|