DQRID : D990419.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/13/1999000004/14/19992359Incorrect
more
Subject:
NSA/MFR10M/C1 - Instrument rotated skyward to check calibration
DataStreams:nsamfr10mC1.a0, nsamfr10mC1.b1
Description:
For purposes of checking the calibration stability, the GNDMFR at the NSA Barrow was 
turned upside down for two days (19990413-19990414).  So instead of looking at the ground, 
this instrument was pointed toward the zenith. 

When the GNDMFR was turned outside down and compared with the MFRSR data, very good 
agreement between the two data streams was observed.  This indicates that both instruments' 
calibrations are remaining stable.
Suggestions: 
Do not use data on these days
Measurements:nsamfr10mC1.b1:
  • callang_flags
  • up_hemisp_narrowband_filter3
  • up_hemisp_narrowband_filter2
  • up_hemisp_narrowband_filter1
  • cosine_solar_zenith_angle
  • up_hemisp_broadband
  • up_hemisp_narrowband_filter6
  • up_hemisp_narrowband_filter5
  • up_hemisp_narrowband_filter4
more
nsamfr10mC1.a0:
  • up_hemisp_narrowband_filter6
  • up_hemisp_narrowband_filter4
  • up_hemisp_narrowband_filter5
  • up_hemisp_broadband
  • up_hemisp_narrowband_filter2
  • up_hemisp_narrowband_filter3
  • up_hemisp_narrowband_filter1
  • callang_flags