Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/13/1999 | 1420 | 10/14/1999 | 2230 | Incorrect |
Subject: | SGP/THWAPS - THWAPS pressure too high by factor of 1.1 |
DataStreams: | DsgpthwapsC1.b1, DsgpthwapsC1.00, DsgpthwapsC1.a1 |
Description: | When attempting to reprogram the C1 THWAPS Campbell Scientific data logger to correct the problem (previously reported) with RH, Site Ops mistakenly changed the scaling of the pressure sensor instead. Pressure data in the ingested file is 1.1 too high. The RH is 1.1 too low (subject of a different DQR). |
Suggestions: | |
Measurements: | DsgpthwapsC1.b1: more
DsgpthwapsC1.00: more DsgpthwapsC1.a1: more |