Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/23/1999 | 0000 | 04/01/2000 | 2359 | Suspect |
Subject: | TWP/SMET/C1 - Incorrect calibration for ARCS1 Wind speed |
DataStreams: | twpsmetC1.00, twpsmet60sC1.a1, twpsmet60sC1.b1, twpsmet60sC1.00 |
Description: | On 23 OCT 99, during Reset 8, an incorrect slope of was entered for the upper, windspeed1, wind speed sensor resulting in a 1% overestimate of the wind speed. The correct calibration was entered on 1 APR 2000 during RESET 10. |
Suggestions: | The data during this period should be reprocessed to reduce the upper wind speeds by 1%. |
Measurements: | twpsmetC1.00: more
twpsmet60sC1.b1: more twpsmet60sC1.a1: more twpsmet60sC1.00: more |