Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/17/2001 | 0300 | 04/30/2002 | 2359 | Does not affect quality |
Subject: | TWP/SKYRAD/C2 - Reprocess: PIR identified as unshaded is really shaded |
DataStreams: | twpskyrad60sC2.b1 |
Description: | On 17Sep01 the Global PIR was moved from the unshaded position on the Skyrad stand to the new 'shaded' spot on the new K&Z tracker. It is still using the same Logger channel as it did as an unshaded PIR even though it is now shaded. |
Suggestions: | The unshaded PIR is now a shaded PIR. Disregard the data description in the netcdf header |
Measurements: | twpskyrad60sC2.b1: more
|