DQRID : D031001.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/01/2002000012/12/20032359Suspect
more
Subject:
NSA/WSI/C1/C2 - Internal dome fogging/contamination
DataStreams:nsawsifullradianceC2.b1, nsawsipatchsummaryC2.b1, nsawsipatchsummaryC1.b1,
nsawsifullradianceC1.b1
Description:
Outgassing from an internal component condenses over time on the optical dome, resulting 
in a slight fogging which is periodically cleaned off.  All radiance-based files will be 
minimally affected in absolute value but will be degraded in spatial resolution.  No cloud 
retrieval file in the archive will be affected because the WSI QC process removes bad 
files.
Suggestions: 
Maintain vigilance.  For highly detailed use of the data, contact the mentor.
Measurements:nsawsipatchsummaryC2.b1:
  • patch_sd
  • patches
  • patch_avg
more
nsawsifullradianceC2.b1:
  • cal_image
more
nsawsipatchsummaryC1.b1:
  • patch_sd
  • patch_avg
  • patches
more
nsawsifullradianceC1.b1:
  • cal_image
more

Close this window