Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/22/2003 | 0447 | 10/12/2003 | 2359 | Suspect |
Subject: | SGP/SMOS/E4 - Sensor failure--RH data bad |
DataStreams: | sgp1smosE4.b1, sgp1smosE4.a1, sgp1440smosE4.a1, sgp30smosE4.b1, sgp30smosE4.a1 |
Description: | RH readings are questionable during high moisture conditions. Readings frequently are above tolerance (104%) during saturated conditions. During non-saturated conditions the RH% readings are fine. |
Suggestions: | |
Measurements: | sgp30smosE4.a1: more
sgp1smosE4.a1: more sgp1440smosE4.a1: more sgp30smosE4.b1: more sgp1smosE4.b1: more |