Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/03/2003 | 1700 | 03/13/2003 | 2130 | Incorrect |
Subject: | TWP/QCRAD/C1 - Upwelling flux incorrect |
DataStreams: | twpqcrad1longC1.s1, twpqcrad1longC1.c1, twpqcrad1longC1.s2, twpqcrad1longC1.c2 |
Description: | An incorrect datalogger configuration made upwelling fluxes incorrect during this time period. Described in instrument DQR D030317.3: The configuration for the SMET was inadvertently loaded into the GNDRAD at Manus at 03/03/2003 17:00 GMT. This error was found and corrected on 03/13/2003 at 21:30 GMT. Because the wrong channels were addressed with incorrect sampling coefficients, it would be very difficult (and in some cases impossible) to reconstruct the data for this week and 3 day period. |
Suggestions: | Don't use upwelling fluxes during this time period. |
Measurements: | twpqcrad1longC1.c1: more
twpqcrad1longC1.s1: more twpqcrad1longC1.s2: more twpqcrad1longC1.c2: more |