Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/21/2003 | 1636 | 01/27/2004 | 2208 | Incorrect |
Subject: | SGP/AOS/C1 - Ozone Instrument removed for repair |
DataStreams: | sgpaosauxC1.a0, sgpaosC1.a0 |
Description: | The ozone instrument was removed for calibration on 2003 11/21. It was shipped to Boulder and was cleaned and calibrated. It was sent back to SGP on 2003 12/02, but was damaged in transit by FedEx. The instrument was reinstalled on 2003 12/03 but was found to be defective. Large spikes were observed in the data. Afetr a few days the instrument was removed and shipped to a vendor for repair. The ozone monitor was returned to service after repair at 2004 01/27 (2208 GMT). |
Suggestions: | Do not use ozone data from this period. |
Measurements: | sgpaosauxC1.a0: more
sgpaosC1.a0: more |