DQRID : D040206.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/13/2003000001/28/20042030Suspect
more
Subject:
SGP/EBBR/E7 - Pressure Sensor Data Incorret
DataStreams:sgp15ebbrE7.b1, sgp5ebbrE7.b1, sgp30ebbrE7.b1
Description:
The E7 pressure sensor suffered
from spikes in pressure, which were related to 
increases in voltage during the day as the 
solar panels charged the battery.  These spikes
were not of a magnitude that would affect the 
30 minute surface fluxes.

Replacement of the pressure sensor solved the problem.
Suggestions: 
Measurements:sgp5ebbrE7.b1:
  • atmos_pressure
more
sgp15ebbrE7.b1:
  • atmos_pressure_mv
more
sgp30ebbrE7.b1:
  • atmos_pressure
more

Close this window