Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/02/2003 | 0000 | 01/23/2004 | 2359 | Suspect |
Subject: | TWP/AERI/C2 - Data Loss due to Communication Problem |
DataStreams: | twpaerisummaryC2.b1, twpaeriengineerC2.b1, twpaerirawC2.00, twpaerich1C2.b1, twpaerich2C2.b1 |
Description: | The spectrometer was periodically unable to transfer data to the computer. When this occurred, the data was deleted and lost as new data filled the spectrometer's memory. All existing data in this time period is good. However, much data is missing. For example, half of channel 1 and three-quarters of channel 2 may be missing. The DSP card was failing intermittantly and the problem was resolved when the card was replaced. |
Suggestions: | Users can safely use any data that exists but missing data can NOT be retrieved |
Measurements: | twpaerisummaryC2.b1: more
twpaerich1C2.b1: more twpaerich2C2.b1: more twpaeriengineerC2.b1: more twpaerirawC2.00: more |