DQRID : D040210.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/20/2004223502/17/20042243Does not affect quality
more
Subject:
SGP/MWR/B6 - Heater problem
DataStreams:sgpmwrtipB6.a1, sgpmwrlosB6.b1, sgpmwrlosB6.a1
Description:
When the heater/blower was replaced, the wet_window flag was constantly set high because 
the "moisture threshold voltage" value in the configuration file was not changed. The 
value was corrected for the new heater and the constant flagging stopped.
Suggestions: 
Measurements:sgpmwrtipB6.a1:
  • wet_window
more
sgpmwrlosB6.b1:
  • wet_window
more
sgpmwrlosB6.a1:
  • wet_window
more