Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/20/2004 | 2235 | 02/17/2004 | 2243 | Does not affect quality |
Subject: | SGP/MWR/B6 - Heater problem |
DataStreams: | sgpmwrtipB6.a1, sgpmwrlosB6.b1, sgpmwrlosB6.a1 |
Description: | When the heater/blower was replaced, the wet_window flag was constantly set high because the "moisture threshold voltage" value in the configuration file was not changed. The value was corrected for the new heater and the constant flagging stopped. |
Suggestions: | |
Measurements: | sgpmwrtipB6.a1: more
sgpmwrlosB6.b1: more sgpmwrlosB6.a1: more |