TWP/MWR/C1 - wrong azimuth
DataStreams:twpmwrlosC1.b1
Description:
The MWR was initially installed at an azimuth angle defined as 180 degrees but the value 
in the configuration file was not changed from the default of 0 degrees. In examining 
photos taken during the installation of the AWS tower, I noticed that the MWR was rotated 
opposite the normal orientation. The value in the configuration file was changed to reflect 
the actual azimuth of the instrument.
Suggestions: 
The scientifically relevant data are not affected.
Measurements:twpmwrlosC1.b1:
  • actaz
more