Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/18/2004 | 0830 | 02/24/2004 | 0830 | Incorrect |
Subject: | NSA/MPL/C1 - Instrument problem |
DataStreams: | nsamplC1.a1, nsamplC1.00, nsamplpsC1.a0 |
Description: | On 2/18, the energy monitor on the MPL dropped from 11 microjoules per second to 2 microjoules per second. This severely affected data quality. This appears to be have been an intermittent cable problem for the laser Q-Switch. A new cable was provided and installed on 2-24. |
Suggestions: | No work around is available. |
Measurements: | nsamplpsC1.a0: more
nsamplC1.a1: more nsamplC1.00: more |