DQRID : D040425.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/21/2004123204/21/20041248Incorrect
more
Subject:
TWP/SONDE/C2 - Bad RH values aloft
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit on this sonde failed about 24 minutes into 
the flight.  The RH values look reasonable, though somewhat noisy, from the surface to 
approximately 6.9 km (439 hPa) at which point they begin to oscillate between 0 and 100 %.  
This behavior stops when the sonde reaches approximately 11.5km (232 hPa) when the air 
temperature drops below -40 degC.  The oscillation stops at that point, though the RH 
remains higher than might be expected even under supersaturated conditions.
Suggestions: 
I suggest discarding all RH data collected above 6 km.
Measurements:twpsondewnpnC2.b1:
  • rh
  • dp
more