DQRID : D040627.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/21/2004001006/21/20040011Incorrect
more
Subject:
TWP/SONDE/C1  - Bad RH in short interval aloft
DataStreams:twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit on this sonde failed about 31 minutes into 
the flight.  The RH values look reasonable from the surface to approximately 8.25 km at 
which point they begin to oscillate between 0 and 40%.  This behavior stops when the sonde 
reaches approximately 11.0km when the air temperature drops below -40 degC.  The 
oscillation stops at that point, though the RH remains higher than might be expected even under 
supersaturated conditions.
Suggestions: 
I suggest discarding all RH data collected between 8.25-11.0 km.
Measurements:twpsondewnpnC1.b1:
  • dp
  • rh
more