Description:
Water entered the Net Radiometer through a cracked upper dome, causing the
net radiation values to be too large.  As a result, latent and sensible heat flux values 
are also incorrect.
Suggestions: 
Measurements:sgp5ebbrE22.b1:
  • net_radiation
more
sgp30ebbrE22.b1:
  • sensible_heat_flux
  • latent_heat_flux
  • net_radiation