Subject: | TWP/SKYRAD/C1 - UVB Data questionable/incorrect due to uncontrolled sensor temperature |
DataStreams: | twpskyrad60sC1.b1, twpskyrad20sC1.a1
|
Description: | The UVB was not temperature controlled to 25C as required
to ensure accurate data. UVB data should ONLY be used after
checking both the bit_data and twpskyrad20sC1.a1 uvb_temp
or inst_down_uvb_temp. UVB data where the UVB temperature
was outside the range of 24-26C should not be used. |
Suggestions: | Use all UVB data from Manus with caution, checking the inst_down_uvb_temp info in the
twpskyrad20sC1.a1 data stream to ensure the temperature was maintained within the valid
range of 24-26C. |
Measurements: | twpskyrad60sC1.b1: - down_uvb_std
- down_uvb_min
- down_uvb_max
- down_uvb
more
twpskyrad20sC1.a1: - inst_down_uvb_irradiance
- inst_down_uvb_temp
more
|