Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/03/2004 | 0307 | 08/17/2004 | 2138 | Incorrect |
Subject: | TWP/SMET/C1 - Incorrect rainrate equation used |
DataStreams: | twpsmet60sC1.b1 |
Description: | There are two different Optical Raingages in use at the TWP sites. The incorrect equation was used at this site. The incorrect equation is (25*(V^1.87)) - 0.15. The correct equation is (20*(V^2)) - 0.05. The correct equation was installed. |
Suggestions: | Solve incorrect equation to find the voltage value. Apply correct equation to solved voltage value to obtain correct reading. |
Measurements: | twpsmet60sC1.b1: more
|