DQRID : D040904.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/24/2004000009/02/20042350Does not affect quality
more
Subject:
TWP/SKYRAD/C3 -  Reprocessed and Corrected: Incorrect NIP cal coefficient
DataStreams:twpskyrad60sC3.b1
Description:
DATA REPROCESSED and CORRECTED: 

During installation of the new Campbell datalogger system and associated program, the NIP 
(S/N 33551E6) was assigned an incorrect calibration coefficient. A coefficient of 116.01 
W/m2/mV was applied when it should have been 124.378 W/m2/mV. The correct calibration 
coefficient has now been applied.

The data were archived April 2017. Please retrieve the reprocessed data for use in your 
analysis.

You may find these data using the ARM Discovery tool at the following link:
http://www.archive.arm.gov/discovery/#v/results/s/finst::skyrad/fsite::twp/ffac::twp.C3 


ORIGINAL DQR CATEGORY: Red( Incorrect)

REPROCESSED and CORRECTED DATA ARCHIVED:2017-04

REPROCESS IDENTIFICATION NUMBER: RID 184

---------------------------------------------------
--- ORIGINAL DQR DESCRIPTION ---
---------------------------------------------------


During our site visit comparison at Darwin we observed that the skyrad NIP was reading 
about 50 w/m2 lower than the comparison cavity radfiometer and the replacement NIP on the 
comparison stand. We found that the calibration coefficient for the NIP (responsivity at 
45o)was put in as 8.64X10-6. The NIP was new and had not been calibrated at SGP. The 
factory number was 8.04X10-6. We corrected the calibration coefficient and the instruments 
agreed to better than 5 w/m2.