Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/31/2004 | 1300 | 09/09/2004 | 1500 | Incorrect |
Subject: | SGP/EBBR/E2 - Net Radiometer Dome Broken, Reducing Net Radiation Measurement, Fluxes |
DataStreams: | sgp5ebbrE2.b1, sgp30ebbrE2.b1, sgp15ebbrE2.b1 |
Description: | The upper dome of the net radiometer was broken. Normally this alone does not affect the net radiation measurement. However, in this case the breakage must have been more than just a bird claw hole or a crack, resulting in obstruction of the top sensing surface of the net radiometer and greatly reduced net radiation measurements. Sensible and latent heat fluxes were reduced as a result. |
Suggestions: | |
Measurements: | sgp30ebbrE2.b1: more
sgp5ebbrE2.b1: more sgp15ebbrE2.b1: more |