Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/09/2004 | 0000 | 10/15/2004 | 2359 | Does not affect quality |
Subject: | NSA/GNDRAD/C1 - Frosting/Condensation |
DataStreams: | nsagndradC1.00, nsagndrad20sC1.a1, nsagndrad60sC1.b1 |
Description: | The GNDRAD PIR and IRT comparison was showing the IRT 2-4 W/m^2 greater than the PIR. Suspected snow or frost on the instrument(s). Corrective maintenance performed 10/15/04 at 21:30 UT. Optics of both instruments were cleaned. Upwelling longwave measurement agreement restored (+/- 5 W/^2). Although the relative performance (agreement) of the two instruments improved after corrective maintenance (+/- 2 W/m^2), no correction is suggested because the agreement remains with the estimated measurement uncertainty of +/- 4 W/m^2 (IRT) and +/-10 W/m^2 (Inverted PIR) |
Suggestions: | |
Measurements: | nsagndrad60sC1.b1: more
nsagndradC1.00: more nsagndrad20sC1.a1: more |