Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/30/2004 | 1835 | 10/13/2004 | 2118 | Suspect |
Subject: | SGP/MWR/C1 - thermal instability |
DataStreams: | sgpmwrlosC1.b1, sgpmwrtipC1.a1, sgpmwrlosC1.a1 |
Description: | The analog board was replaced with a spare (D041001.3) with a reference temperature that was set too low (306 K) so that during periods of high ambient temperature, the instrument became thermally unstable. The problem was corrected when the temperature setting was increased (to 311 K). |
Suggestions: | The data are suspect when the air temperature, tkair, exceeds 297 K. |
Measurements: | sgpmwrlosC1.a1: more
sgpmwrtipC1.a1: more sgpmwrlosC1.b1: more |