DQRID : D041021.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/02/2004000011/05/20042359Suspect
more
Subject:
SGP/NFOV/C1 - Condensation on window
DataStreams:sgpnfovC1.b1, sgpnfovC1.a0
Description:
The single-channel NFOV had a condensation problem.  Water made its way into the interior 
of the instrument and condensed on some of the important optical components of the system.

The condensation resulted in the signal strength of the 1-NFOV being significantly less 
than that of the 2-NFOV.
Suggestions: 
Compare data with the 2-channel NFOV to determine on which days the problem occurred.
Measurements:sgpnfovC1.b1:
  • radiance_temp_cor
  • raw_counts
  • callang_flags
  • radiance
more
sgpnfovC1.a0:
  • radiance
  • ambient_temp
  • callang_flags
  • raw_counts
more

Close this window