DQRID : D041026.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/28/2004140008/27/20041540Missing
more
Subject:
SGP/AOS/C1 - Condensation Particle Counter failure
DataStreams:sgpaosauxC1.a0, sgpaosC1.a0
Description:
The CPC signal declined to a low value. The instrument was shipped back to the mentor, 
cleaned and repaired. The instrument inlet was clogged.
Suggestions: 
Data are missing and unrecoverable.
Measurements:sgpaosauxC1.a0:
  • CPCCondT
  • CPCSaturT
  • CPCDeltT
more
sgpaosC1.a0:
  • CPCPartConc
more