Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/24/2004 | 0000 | 02/18/2004 | 1800 | Missing |
Subject: | SGP/AOS/C1 - Wet nephelometer problems |
DataStreams: | sgpaosC1.a0, sgpaosauxC1.a0 |
Description: | The humidifier tube was coated with mold and had to be replaced. This affects the wet nepheleometer measurements and the calculations of the aerosol hygroscopic growth. The humidifer was cleaned and replaced. A filter was added to the water circulation to minimize mold. |
Suggestions: | |
Measurements: | sgpaosauxC1.a0: more
sgpaosC1.a0: more |