DQRID : D041102.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/13/2004060010/18/20042300Does not affect quality
more
Subject:
TWP/AERI/C2 - Bias incomparison to SKYRAD-IRT
DataStreams:twpaerich1C2.b1
Description:
During the time period specified, the IRT measured higher sky temperatures
than the AERI.  While it is not possible to determine which instrument is
biased, the AERI is calibrated between each sample and thus its data has
a high degree of confidence.
Suggestions: 
Measurements:twpaerich1C2.b1:
  • mean_rad
more