Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/01/2002 | 0000 | 09/17/2004 | 1745 | Suspect |
Subject: | SGP/RL/C1 - Decreased sensitivity |
DataStreams: | sgp10rlprofbe1turnC1.c0, sgp10rlprofdep1turnC1.c1, sgp10rlprofasr1ferrC1.c1, sgprlC1.00, sgp10rlprofbe1turnC1.c1, sgprlC1.a0, sgp10rlprofext1ferrC1.c1, sgp10rlprofmr1turnC1.c1 |
Description: | The sensitivity of the Raman Lidar degraded significantly beginning in late 2001. Between May and September 2004, the electronics were upgraded and the telescope was refurbished resulting in sensitivity better than when the instrument system was first installed. A detailed technical report describing the effects of the signal degradation and the steps taken to correct the problem is being written. In the interim, contact Dave Turner (dave.turner@pnl.gov) for more information. |
Suggestions: | Data should be used with caution. Contact Turner for more information. |
Measurements: | sgp10rlprofext1ferrC1.c1: more
sgp10rlprofasr1ferrC1.c1: more sgp10rlprofmr1turnC1.c1: more sgprlC1.00: more sgprlC1.a0: more sgp10rlprofdep1turnC1.c1: more sgp10rlprofbe1turnC1.c0: more sgp10rlprofbe1turnC1.c1: more |