DQRID : D050225.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/05/2004150012/03/20041700Does not affect quality
more
Subject:
SGP/BRS/C1 - Reprocessed and Corrected: Incorrect Calibration Coefficients
DataStreams:sgpbrsC1.00, sgpbrs20sC1.a0, sgpbrsC1.b1
Description:
DATA REPROCESSED and CORRECTED: 


Lightning strike on or near 6/2/2004 took the BRS NIP (31747E6) out of service, on or near 
6/15 the NIP was replaced and the datalogger program was changed to reflect the new NIP 
(31343E6). On 8/5/2004 the DIR (Downwelling longwave) instrument was changed out and a 
new program was loaded to reflect a new DIR. The new program contained the original 
installed shortwave instrument calibration coefficients including NIP 31747E6. Thus from 8/5 to 
when the shortwave instruments were changed out on 12/3 the wrong s/n and calibration 
factor (i.e. 31747E6) was applied to the actual NIP (31343E6) installed at BRS C1. The 
calibration coefficients now correctly match NIP 31343E6.

The data were archived in May 2017. Please retrieve the reprocessed data for use in your 
analysis.

You may find these data using the ARM Discovery tool at the following link:
http://www.archive.arm.gov/discovery/#v/results/s/finst::brs/fsite::sgp/ffac::sgp.C1 

ORIGINAL DQR CATEGORY: Red( Incorrect)

REPROCESSED and CORRECTED DATA ARCHIVED:2017-05

REPROCESS IDENTIFICATION NUMBER: RID 190

---------------------------------------------------
--- ORIGINAL DQR DESCRIPTION ---
---------------------------------------------------


Lightening strike on or near 6/2/2004 took the BRS NIP(31747E6) out of service, on or near 
6/15 the NIP was replaced and the datalogger program was changed to reflect the new 
NIP(31343E6). On 8/5/2004 the DIR (Downwelling longwave) instrument was changed out and a new 
program was loaded to reflect a new DIR.  The new program contained the original 
installed shortwave instrument calibration coefficients including NIP 31747E6. Thus from 8/5 to 
when the shortwave instruments were changed out on 12/3 the wrong s/n and calibration 
factor (i.e. 31747E6) was applied to the actual NIP(31343E6)installed at BRS C1.
Suggestions: 
Measurements:sgpbrs20sC1.a0:
  • inst_direct_normal
more
sgpbrsC1.00:
  • Raw data stream - documentation not supported
more
sgpbrsC1.b1:
  • short_direct_normal_std
  • short_direct_normal_min
  • short_direct_normal_max
  • short_direct_normal
more