Subject:
TWP/QCRAD/C3 - MFRSR shading problem
DataStreams:twpqcrad1longC3.s1, twpqcrad1longC3.c1, twpqcrad1longC3.s2, twpqcrad1longC3.c2
Description:
MFRSR diffuse and direct fluxes incorrect during this time because instrument was not 
shading properly.  Some tests on best estimate global shortwave cannot be applied.  BEGSW 
could be incorrect if MFRSR data used in tests or fits.

Documented in instrument DQR D030317.2, D050106.2
Suggestions: 
Do not use BestEstimate_down_short_hemisp if source_BestEstimate_down_short_hemisp = 

-2; Do not use diffuse/direct MFRSR fields in QCRAD for this time period.
Measurements:twpqcrad1longC3.s2: