DQRID : D050117.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/23/2004231501/25/20051800Incorrect
more
Subject:
SGP/SIRS/E9 - Unknown -failed PSP ventilator fan
DataStreams:sgpsirs20sE9.a0, sgpsirsE9.b1, sgpsirsE9.00
Description:
Downwelling shortwave hemispheric (DSH) readings appear to have a slight to moderate 
negative offset from derived DSH values.  At night, this negative offset causes qc flags 
(since values are below the minimum threshold of -10 W/m^2).  Differences between derived and 
actual DSH values are sometimes on the order of 50 W/m^2 due to this offset. 
The offset was discovered to be due to a failed ventilator fan which is used to clear the 
dome of the PSP, the instrument which measures the DSH.  The ventilator likely seized or 
was intermittent and heated the PSP base giving an erroneous DSH reading.  Field techs 
replaced the fan on 01/25/2005 1745GMT.  The DSH has been reading normally since that time.
Suggestions: 
Disregard this data.