DQRID : D050303.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/10/2005162003/10/20051600Does not affect quality
more
Subject:
SGP/SIRS/E2 - Reprocessed: Incorrect DIR calibration coefficients
DataStreams:sgpsirs20sE2.a0, sgpsirsE2.00, sgpsirsE2.b1
Description:
During the DIR/UIR instrument changeout on 2/10/2005 the new datalogger program contained 
incorrect calibration coefficients for the DIR instrument. The program was corrected on 
3/10/2005.

These data were reprocessed and archived in September 2006.
Suggestions: 
Measurements:sgpsirs20sE2.a0:
  • inst_down_long_shaded_case_resist
  • inst_down_long_shaded_dome_resist
  • inst_down_long_hemisp_shaded_tp
more
sgpsirsE2.00:
  • Raw data stream - documentation not supported
more
sgpsirsE2.b1:
  • down_long_hemisp_shaded_min
  • down_long_hemisp_shaded_std
  • inst_down_long_shaded_dome_temp
  • inst_down_long_shaded_case_temp
  • down_long_hemisp_shaded_max
  • down_long_netir
  • down_long_hemisp_shaded
more

Close this window