DQRID : D050414.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/12/2005000005/06/20051900Incorrect
more
Subject:
SGP/NFOV/C1 - Head temp heater circuit failed
DataStreams:sgpnfovC1.b1, sgpnfovC1.a0
Description:
The head temp values were erratic.  The instrument was removed from service and found the 
head temp heater circuit failed in the full-on position frying the detectors.  Data are 
not useful.
Suggestions: 
Do not use data during this time period.
Measurements:sgpnfovC1.a0:
  • head_temp
  • ambient_temp
  • tube_temp
  • callang_flags
  • radiance
  • raw_counts
more
sgpnfovC1.b1:
  • callang_flags
  • tube_temp
  • cosine_solar_zenith_angle
  • head_temp
  • radiance
  • radiance_temp_cor
  • raw_counts
  • ambient_temp
more

Close this window