DQRID : D050421.14
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/03/2004000002/23/20052300Incorrect
more
Subject:
SGP/MPL/C1 - Sensor Calibration
DataStreams:sgpmplC1.a1, sgpmplC1.00
Description:
The Energy Monitor records the value of the laser energy. During the time in question, the 
laser energy exceed the maximum value in the calibration table for the energy monitor. 
In other words, the value is pegged. This condition does not affect the backscatter 
signal. But it will prevent users from normalizing the backscatter signal to the laser output.
Suggestions: 
Measurements:sgpmplC1.00:
  • Raw data stream - documentation not supported
more
sgpmplC1.a1:
  • energy_monitor
more

Close this window