Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/25/2005 | 1600 | 04/01/2005 | 2310 | Incorrect |
Subject: | NSA/MPL/C1 - Diode failure |
DataStreams: | nsamplC1.a1, nsamplpsC1.a0 |
Description: | On 3/25, the energy monitor on the MPL dropped from 11 microjoules per second to 2 microjoules per second. We determined the problem was the diode; it reached the end of its lifetime. A replacement was sent. Data quality returned to normal. |
Suggestions: | No work around is available. |
Measurements: | nsamplpsC1.a0: more
nsamplC1.a1: more |