DQRID : D050912.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/09/2005113209/09/20051332Incorrect
more
Subject:
SGP/SONDE/C1 - Temperature/RH incorrect
DataStreams:sgpsondewnpnC1.b1, sgplssondeC1.c1
Description:
The temperature throughout this sounding is too high and as a result the RH values will be 
incorrect as well.  I'm not sure what is wrong here but the operators noted reasonable 
values (18.1 degC, 87%) in the log before launch.  Review of the raw data, however, shows 
the sonde reporting temperatures nearly 20 degC higher.  I suppose it is possible that 
the operator entered into the log the reference temperature from the GC25 ground check set 
rather than the temperature reported by the sonde - I can't think of any other 
explanation - there is no doubt that the temperature sensor on this sonde was bad.
Suggestions: 
No fix possible - ignore the data.