DQRID : D051106.16
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/17/2005163008/27/20051930Incorrect
more
Subject:
SGP/EBBR/E27 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorredt
DataStreams:sgp5ebbrE27.b1, sgp30ebbrE27.b1, sgp15ebbrE27.b1
Description:
Residual Moisture left in the top dome of the net radiometer after the 8/17/05 servicing 
caused the net radiation measurements to be biased low, and resulting in sensible and 
latent heat flux being low also.
Suggestions: 
Measurements:sgp5ebbrE27.b1:
  • net_radiation
more
sgp15ebbrE27.b1:
  • net_radiation_mv
more
sgp30ebbrE27.b1:
  • net_radiation
  • sensible_heat_flux
  • latent_heat_flux
more

Close this window