DQRID : D051121.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/13/2005234604/15/20051732Incorrect
more
Subject:
SGP/SONDE/C1 - Sonde RH values incorrect
DataStreams:sgpsondewnpnC1.b1
Description:
RH values for all sondes launched during this time were incorrect due to a faulty Ground 
Check set which damaged or conditioned the RH sensor improperly.
Suggestions: 
Measurements:sgpsondewnpnC1.b1:
  • rh
  • dp