DQRID : D060630.33
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/20/2004220012/21/20052136Does not affect quality
more
Subject:
TWP/SKYRAD/C3 - Reprocessed: Longwave Calibration error
DataStreams:twpskyrad60sC3.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.

The data collected while the incorrect procedures were in place were reprocessed to remove 
the calibration bias.  Note, this also affected standard deviation, maximum and minimum 
data fields.
Suggestions: 
Reprocessed data were archived 20100211.
Measurements:twpskyrad60sC3.b1:
  • down_long_hemisp2
  • down_long_hemisp1
  • down_long_hemisp1_std
  • down_long_hemisp1_max
  • down_long_hemisp2_min
  • down_long_hemisp2_std
  • down_long_hemisp1_min
  • down_long_hemisp2_max
more

Close this window