DQRID : D040421.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/28/2003020006/21/20060215Suspect
more
Subject:
NSA/SKYRAD/C1 - itermittant IRT bias
DataStreams:nsaskyrad60sC1.b1
Description:
A problem with the SKYRAD IRT that itermittantly produced a gradual increase in the sky 
temperature measurement then a sudden correction, as compared to AERI, appears to have been 
caused by a combination of a failing IR detector, a poor signal cable, a contaminated 
(with condensation or frost) mirror or lens, and low ambient temperatures. The positive 
bias was apparently not consistant in either frequency or duration but on days when the 
unusual behavior is observed, it usually occurred from about 00:00 to 06:00 local time. 

The instrument was repaired and recalibrated and the signal cable replaced.
Suggestions: 
Measurements:nsaskyrad60sC1.b1:
  • sky_ir_temp
  • sky_ir_temp_max
  • sky_ir_temp_std
  • sky_ir_temp_min
more